A comparison for ultrasound energy delivery between the new balanced tip and MiniTip for torsional phaco
(results will display both Free Papers & Poster)
Session Details
Session Title: Cat Surgery Equip/Phaco
Session Date/Time: Monday 15/09/2014 | 14:30-16:00
Paper Time: 14:30
Venue: Boulevard F (Level 1)
First Author: : K.Tjia THE NETHERLANDS
Co Author(s): :
Abstract Details
Purpose:
To compare the Cumulative Dissipated Energy(CDE), between Infiniti Vision System with the 800 micron Minitip and Centurion Vision System with the new Balanced tip
Setting:
Isala Clinics, Zwolle, the Netherlands
Methods:
60 cases for each group were matched for nuclear density. The Infiniti Vision System group Ultrasound(US) settings were maximized at 80% amplitude during Sculpt and Quadrant removal(QR). Vacuum setting during QR was maximum 300mm Hg.
The Centurion Vision System US settings were maximized at 80% during sculpting and 65% during QR. Vacuum was set at a maximum of 550mm Hg.
Results:
Preliminary results show a 40% reduction in cumulative dissipated energy during sculpting and quadrant removal for the Centurion group.
Conclusions:
The new Balanced tips with Centurion Vision System at the employed higher Vacuum setting are significantly more energy efficient compared to Infiniti with Minitips.
Financial Interest:
One or more of the authors... receives consulting fees, retainer, or contract payments from a company producing, developing or supplying the product or procedure presented, One or more of the authors... travel has been funded, fully or partially, by a company producing, developing or supplying the product or procedure presented