Comparison of incision integrity and morphological features following IOL implantation with three injector devices: randomized clinical trial
(results will display both Free Papers & Poster)
Session Details
Session Title: Cataract Surgery/ Equip/ IOLs
Session Date/Time: Sunday 14/09/2014 | 14:30-16:00
Paper Time: 14:42
Venue: Capital Hall B
First Author: : S.Vasavada INDIA
Co Author(s): : V. Vasavada A. Vasavada S. Raj V. Vasavada S. Srivastava
Abstract Details
Purpose:
Evaluate and compare incision enlargement as well as ingress of trypan blue from the ocular surface into the anterior chamber following implantation of a foldable hydrophobic acrylic intraocular lens (IOL) through a 2.2mm clear corneal incision(CCI) using one automated and two manual injector systems
Setting:
Iladevi Cataract & IOL Research Centre, Raghudeep Eye Hospital,
Methods:
Randomized clinical trial in120 eyes(120 patients)undergoing microcoaxial phacoemulsification through a 2.2mm CCI randomized to three groups depending on IOL injector used:Group I(n=40),where automated motorized injector(Autosert®) used;Group II(n=40),where a plunger type injector(Royale)used;Group III(n=40),screw type injector(Monarch III)used.Incision widths were measured before and after implantation of a foldable,single-piece aspheric hydrophobic acrylic IOL using D cartridge.Following IOL implantation and stromal hydration,trypan blue 0.0125% was applied over the conjunctival surface,and ingress compared.Architectural features of CCIs were evaluated using anterior segment OCT(AS-OCT) on postoperative day 1,1week and 1month
Results:
Ongoing study,results of 60 eyes presented,results will be updated.Incision enlargement from end of phacoemulsification to IOL implantation:Group I(n=19):0.07+0.05, Group II(n=18):0.16+0.05,Group III(23):0.22+0.04;difference being statistically significant(P=0.03).Incision enlargement was greatest with Monarch III injector,followed by Royale,and Autosert.Trypan blue concentration into anterior chamber was(log units):GroupI=1.97+0.45,Group II=1.78+0.48,Group III=2.16+0.38;difference not statistically significant(P=0.48).No difference in descemet’s detachment,incision gaping and loss of coaptation between groups at any followup(P>0.05)
Conclusions:
All injectors allowed smooth IOL implantation through a 2.2mm incision;incision enlargement was least using automated injector