Official ESCRS | European Society of Cataract & Refractive Surgeons
London 2014 Registration Visa Letters Programme Satellite Meetings Glaucoma Day 2014 Exhibition Hotel Booking Virtual Exhibition Star Alliance
london escrs

Course handouts are now available
Click here


Come to London

video-icon

WATCH to find out why


Site updates:

Programme Updates. Programme Overview and - Video Symposium on Challenging Cases now available.


Posters

Search Abstracts by author or title
(results will display both Free Papers & Poster)

Numerical approach for interpreting non-contact tonometer results in terms of corneal biomechanical behaviour

Poster Details

First Author: M.Ariza SPAIN

Co Author(s):    D. Piñero   J. Zurita   J. Rodríguez   B. Calvo     

Abstract Details



Purpose:

To evaluate by means of finite element modeling the influence of tissue properties of the cornea on its deformation response when an air jet is applied onto its surface and, according to this, to discuss the clinical interpretation of these measurements in terms of corneal biomechanical behaviour

Setting:

Engineering Research Institute of Aragón, University of Zaragoza (Spain)

Methods:

A patient-specific finite element model of the eye was developed for our simulations. Corneal geometry data obtained with the Pentacam system from Oculus in a healthy cornea were used. Specifically, anterior and posterior surface point clouds were exported used. Three materials of different stiffness were simulated, high, intermediate and low. The material with the lowest stiffness simulated a keratoconus cornea. The influence of intraocular pressure (IOP) was studied by considering three pressure levels: 10, 19 and 28 mmHg. The value of 19 mmHg was the real patient's IOP measured with a Goldmann contact tonometer.

Results:

The deformed configuration of the cornea subjected to an air flow pressure as well as stress and strain fields were computed for each material stiffness and IOP. The maximum displacement of apex showed a linear variation with IOP, with the largest apex displacements for the lowest IOP values. In contrast, maximum apex displacement showed a non-linear relation with tissue stiffness. In general, the cornea only affected by IOP is under traction stress (membrane tensional state), but when it is subjected to air-flow pressure, the anterior surface is under compression whereas the posterior surface is under tension, and therefore the cornea experiences bending. This implies that collagen fibres in the anterior surface do not contribute to load bearing.

Conclusions:

Non-contact tonometry devices, such as CorVis ST or Ocular Response Analyzer, mainly evaluates the corneal deformation response (applanation and maximum displacement) as a consequence of the delivery of an air flow pressure over the corneal apex area. This deformation depends on the mechanical properties of corneal tissue, IOP and corneal thickness, and consequently the evaluation separately of the contribution of each factor is not possible. Thus, a cornea with high stiffness and low IOP may show the same deformation response as a cornea with low stiffness and high IOP. In addition, systems based on non-contact tonometry for characterizing corneal biomechanics evaluates the mechanical response of the cornea under bending and, therefore, the membrane behaviour of the cornea is only evaluated, as in a traction or inflation test. These results show that a complete corneal biomechanical characterization would require more than one test for analysing independently the membrane and bending behaviour of the cornea. FINANCIAL INTEREST: NONE

Back to Poster listing