Course handouts are now available
Click here
Come to London
WATCH to find out why
Site updates:
Programme Updates. Programme Overview and - Video Symposium on Challenging Cases now available.
Posters
(results will display both Free Papers & Poster)
Toric AcrySof IQ rotational stability
Poster Details
First Author: V.Rusu ROMANIA
Co Author(s): G. Rusu I. Astefculesei L. Cucu D. Hamati
Abstract Details
Purpose:
The purpose of this study is to evaluate rotational stability of toric ACRYSOF IQ
Setting:
Spitalul sf Spiridon Iasi Romania
Methods:
This is a four years prospective study on eyes that underwent toric intraocular lens implantation after cataract surgery performed by the same surgeon. Keratorefractometry values, best corrected distace visual acuity were measuread at 2 weeks, 1, 3, 6, 12 month follow-up. Toric IOL alignment was evaluated using stil lamp retroilumination photography. Phothos purchased were pocessed by a computer program PHOTOSHOP and the landmarks were made to calculate the deviation from the predetermined axis. Data were storead in a table in Microsft Office Excel 2007.
Results:
postoperative last control results were: mean best corrected visual acuity 0,9 Snellen; mean residual refractive cylinder was 0,124 D; IOL rotation was less than 5 degree in 81,1% and less than 10 degree in 97,1%.
Conclusions:
toric ACRYSOF IQ rotational stability was good with high visual quality results. FINANCIAL INTEREST: NONE