A comparative study of intraocular pressure measurements using Goldmann Applanation Tonometer, I-care pro, and Corvis ST after DSAEK
Session Details
Session Title: Surgical Cornea
Session Date/Time: Tuesday 25/09/2018 | 16:30-18:00
Paper Time: 17:36
Venue: Room A3, Podium 3
First Author: : S.Pirhadi IRAN
Co Author(s): : K. Jadidi S. Bagheri N. Mohammadi
Abstract Details
Purpose:
To compare the intraocular pressure (IOP) measurements using different devices and evaluate the influence of central corneal thickness (CCT) on IOP measurements following DSAEK surgery.
Setting:
Department of Ophthalmology, Baghiatallah University of Medical Sciences, Tehran, Iran.
Methods:
In this cross-sectional study, 22 eyes from 19 patients (mean age: 56.59 years) after DSAEK with clear and compact graft were enrolled. Mean follow-up after the surgery was 2.191±1.852 years. IOP was measured with GAT, I-care pro and Corvis ST (as non-corrected and biomechanical IOP) in post-DSAEK patients and agreement between measurements of three devises were assessed. Corvis ST was used to measure the CCT value. The influence of CCT on IOP measurements was evaluated using spearman test.
Results:
The mean IOP measured by GAT, I-care pro and Corvis as the non-corrected and biomechanical IOP were 12.545±2.064, 11.591±2.465, 12.686±3.967 and 9.714±4.339 respectively. GAT measurement was considered as a gold standard and the only significant difference was observed between GAT and Corvis biomechanical IOP (P<0.05). CCT did not correlate with any of the IOP measurements except Corvis biomechanical IOP (r= -0.518, P=0.01).
Conclusions:
There was a good agreement between GAT, I care and Corvis non-corrected IOP measurements while the Corvis biomechanical IOP measurement was less than the GAT.
Financial Disclosure:
-