Comparison of 7 modern intraocular lens power calculation formulas for a quadrifocal lens
Session Details
Session Title: IOL Power Calculations, Post-LASIK & Extreme Eyes
Session Date/Time: Tuesday 25/09/2018 | 14:00-16:00
Paper Time: 14:18
Venue: Room A2
First Author: : M.Shajari GERMANY
Co Author(s): : C. Kolb K. Petermann M. Böhm M. Herzog N. De Lorenzo T. Kohnen
Abstract Details
Purpose:
To evaluate the accuracy of 7 formulas (Barrett Universal II, Haigis, Hoffer Q,
HillRBF, Holladay 1, SRK/T, and T2) calculating the power of the quadrifocal
AcrySof® IQ PanOptix® intraocular lens (IOL).
Setting:
Department of Ophthalmology, Goethe University, Frankfurt, Germany.
Methods:
The study comprised 75 eyes of 38 patients having cataract surgery with
insertion of a quadrifocal lens (Panoptix, Alcon) over 15 months. Preoperative biometry
measurements were obtained from an IOLMaster 500. Optimized IOL constants were
calculated to reduce the mean refractive prediction error. Differences in mean absolute
errors between the formulas were evaluated.
Results:
The formulas were ranked by the mean absolute refractive prediction error as
follows: HillRBF (0.274 D), Haigis (0.276 D), T2 (0.295 D), Barrett Universal II (0.298 D),
Holladay 1 (0.328 D), SRK/T (0.342 D), and Hoffer Q (0.347 D). The differences in
absolute errors between the formulas were significant (P < 0.0001).
Conclusions:
The HillRBF and Haigis formula performed very well and had the lowest prediction error and are recommended for iol calculation when implanting a multifocal lens.
Financial Disclosure:
-