Official ESCRS | European Society of Cataract & Refractive Surgeons
Vienna 2018 Delegate Registration Programme Exhibition Virtual Exhibition Satellites 2018 Survey

 

escrs app advert

In vitro comparison of UV transmittance of 8 different hydrophobic IOLs

Search Title by author or title

Session Details

Session Title: New IOLs: Material & Preload IOL

Session Date/Time: Monday 24/09/2018 | 08:00-10:30

Paper Time: 09:57

Venue: Room A3, Podium 2

First Author: : Y.Guldenfels FRANCE

Co Author(s): :    C. Chong   L. Saidi   P. Rozot              

Abstract Details

Purpose:

to determine UV protection provided by various intraocular lenses

Setting:

Acrylian Ltd , 22 rue de la Ganzau 67000 STRASBOURG FRANCE. Clinique RHENA ,10 rue François Epailly 67000 STRASBOURG FRANCE

Methods:

UV transmittance was measured with a KONICA MINOLTA CM5 Spectrophotometer for 8 different IOLs with same power ; transmission was measured from 340 to 740 nm every 10 nm.

Results:

4 different groups of IOLs can e described with different levels of UV protection:first group (Envista) with very low protection : cut off < 370nm; Low protection group with cut off between 370 and 390 nm (Technis, Lucia, Artis, Focusforce); Good protection group 390-400 nm (Eyeceeone); Complete protection cut off between 400 410 nm; Complete protection with color perception modification cut off > 400nm (SN60 &Vivinex)

Conclusions:

UV protection provided by IOLs on the market is different for Esch brand. Knowing these differences can help surgeon to choose IOL depending on patient retinal status.

Financial Disclosure:

... is employed by a for-profit company with an interest in the subject of the presentation, ... travel has been funded, fully or partially, by a company producing, developing or supplying the product or procedure presented