New option of marking the corneal incision and lens axis during femto cataract surgery (LDV Z8, Ziemer) and toric IOL implantation
Session Details
Session Title: Toric IOL Calculations & Alignment
Session Date/Time: Sunday 23/09/2018 | 08:00-10:00
Paper Time: 09:38
Venue: Room A5
First Author: : T.Shimoda BRAZIL
Co Author(s): : G. Shimoda G. Shimoda
Abstract Details
Purpose:
To mark the incision and IOL axis in the cornea with femto LDV Z8 and check the marks in the cornea at slipt lamp immediately after the surgery.
Setting:
Centro de Microcirurgia Ocular Atibaia
Methods:
32 eyes of patient with cataract and cornea astigmatism, submitted to femtocataract surgery with femto LDV Z8, using three marks.
The first was incision axis (2.4 mm x 12.0 mm), second, IOL axis (1.0 mm x 4.00 35% of the cornea), both used 1.200 NJ of energy and both axis given by the calculator.Third mark was 180º of the IOL axis(Second)
The marks were checked at slept lamp immediately after the surgery.
Results:
6 eyes - the error were between 6º - 10º
26 eyes - the error were between 0 -5º
The marks disappeared in 24 hs.
Conclusions:
Mark the axis in the cornea with femto LDV Z8 showed security and precise to implant the toric lens.
Financial Disclosure:
-