Session Title: Phakic IOL Implantation II
Session Date/Time: Wednesday 09/10/2013 | 08:00-10:30
Paper Time: 08:00
Venue: E104-105 (First Floor)
First Author: : J.Cazal SPAIN
Co Author(s): : F. Sanchez Fabo
Purpose:
To evaluate the safety, efficacy, predictability, stability, and complications of the foldable Artiflex Toric iris-fixated intraocular lens and Artisan Toric (Ophtec BV, Groningen, The Netherlands) for the correction of myopia and astigmatism in phakic eyes
Setting:
Centro Oftalmológico del Bages, Manresa,Spain
Methods:
All eyes underwent implantation of the foldable iris-fixated Artiflex phakic intraocular lens (PIOL) with an optic zone of 6 mm. The follow-up was 3 years. Phakic intraocular lenses were implanted in powers ranging from -6.0 to -14.0 D.
Results:
After 3 years, a UCVA of 20/40 or better was observed in 96.2% of eyes. The BSCVA was 20/40 or better in all eyes. A gain of 1 line or more of BSCVA was found in 40.9% of eyes, and none lost 2 or more lines. The mean endothelial cell change was -0.05%, 1.89%, and -2.7% at 6 months, 1 year, and 2 years,respectively. Chronic inflammation was not detected in any eye by slit-lamp biomicroscopy, nor did any eye develop angle closure or glaucoma.
Conclusions:
After 3 years of follow-up results suggest that the Artisan/Artiflex lens is an accurate and safe method for the correction of myopia and astigmatism. Surgical skill is important but could be performed by novel surgeons. Longer-term data are needed to assess the impact of the lens on the endothelium, the crystalline lens, and the iris
Financial Interest:
NONE