First Author: F.Orucoglu TURKEY
Co Author(s): S. Aliyeva
Purpose:
The purpose of this study was to check for five different methods of assessment of astigmatism and axis with auto refractometer, auto keratometer, topography, wavefront and manifest refraction
Setting:
Kudret Eye Hospital
Methods:
The study included 217 eyes of 115 patients who were evaluated for excimer laser procedures. Auto refractometer, auto keratometer, manifest refractive, topographic and wave-front data was evaluated.
Results:
Manifest, auto refractive, keratometric, topographic and wavefront cylinder were -0.97 D, -0.99 D, -1.27 D, -1.21 D, -1.07 D respectively. Auto refractometer and manifest cylinder were not statistically different (p=0. 291). As well keratometric and topographic cylinders were statistically similar (p=0.112). Auto refractive, Auto keratometric, topographic and ,wavefront , axis were 108.4, 96.1, 95.0, 91.6 degrees respectively. Manifest axis were 112.2 degree. Manifest and auto refractometer axis were statistically similar (p=0.193). Keratometric, topographic and wavefront axis were statistically similar (p = 0.570 to 0.825).
Conclusions:
Manifest refraction showed similarity with auto refractometer measurements in normal ametropic eyes. Wavefront axis showed similarity with keratometry and topography. FINANCIAL DISCLOSURE?: No