Evaluation of an electronic toric marker compared with slitlamp marking for cataract surgery
Poster Details
First Author: M. Orts Vila SPAIN
Co Author(s): A. Fernando J. Hueso Abancens
Abstract Details
Purpose:
To compare accuracy of an electronic toric device in marking the axis of astigmatism before cataract surgery compared with manual slitlamp marking.
Setting:
Ophthalmology, Hospital General Universitario de San Juan, Alicante, Spain
Methods:
This prospective clinical study included 39 eyes undergoing cataract surgery with coexisting corneal astigmatism. Eyes were randomly assigned to 2 groups, main incision was made in the steepest meridian. In group A, the steepest axis was marked with an electronic device (Electronic Toric One Step Marker, Asico) held frontally. In group B the horizontal meridian was marked at the slitlamp. Intraoperativelly, the steep meridian was marked using a second device. The difference from the reference axis was calculated using a very narrow slit aligned with the triangular internal mark of the incision done with the point of the knife.
Results:
The electronic toric marker (A group) had a difference of 3.58º (standard deviation 1.4º) from the reference axis. Eyes marked manually with the slit lamp (B Group) had a difference of 7.12 (standard deviation 3.72º) (p<0.001).
Conclusions:
Astigmatism axis marking was more precise with the Electronic One Step Marker. Additionally, this electronic marker was, in our hands, easier and faster to use than manual slitlamp marking.
Financial Disclosure:
NONE