Official ESCRS | European Society of Cataract & Refractive Surgeons
Copenhagen 2016 Registration Programme Exhibitor Information Virtual Exhibition Satellite Meetings Glaucoma Day 2016 Hotel Star Alliance
title

10 - 14 Sept. 2016, Bella Center, Copenhagen, Denmark

This Meeting has been awarded 27 CME credits

 

escrs app advert yo advert

Comparison of the optical quality of different trifocal intraocular lenses using an optical bench

Search Title by author or title

Session Details

Session Title: Pseudophakic IOLs: Multifocal I

Session Date/Time: Monday 12/09/2016 | 08:00-10:30

Paper Time: 09:00

Venue: Auditorium A

First Author: : R.Khoramnia GERMANY

Co Author(s): :    S. Liebing   T. Tandogan   C. Choi   G. Auffarth           

Abstract Details

Purpose:

To compare different trifocal intraocular lenses (IOLs) at the optical bench regarding their optical quality.

Setting:

David J Apple International Laboratory for Ocular Pathology, University Eye Hospital Heidelberg, Heidelberg, Germany

Methods:

We analyzed the FineVision (PhysIOL, Belgium), the AT LISA tri 839MP (Zeiss, Germany) and the AcrySofIQ PanOptix (Alcon, USA) with a power of 21D for the distance using the OptiSpheric optical bench (Trioptics, Germany). The addition for the near and intermediate distances are as follows: +3.5D/+1.75D (FineVision), +3.33D/+1.66D (AT LISA) and +3.2D/+2.2D (PanOptix). We evaluated the modulation transfer function (MTF) at a spatial frequency of 50 lp/mm and the Strehl ratio using a 3 mm and 4.5 mm aperture.

Results:

The MTF at 50lp/mm (FineVision/AT Lisa/PanOptix) at the far focus was 0.385/0.397/0.401 (3mm aperture) and 0.488/0.285/0.435 (4.5mm aperture). At the intermediate focus, the MTF was 0.160/0.155/0.155 (3mm aperture) and 0.093/0.114/0.140 (4.5mm aperture). The MTF at the near focus was 0.218/0.184/0.402 (3mm aperture) and 0.187/0.185/0.177 (4.5mm aperture). The Strehl ratio was 0.347/0.294/0.368 (3mm aperture) and 0.212/0.146/0.297 (4.5mm aperture) at the far focus. At intermediate distances, the Strehl ratio was 0.177/0.188/0.176 (3mm aperture) and 0.088/0.092/0.126 (4.5mm aperture). The Strehl ratio was 0.296/0.281/0.456 (3mm aperture) and 0.151/0.164/0.183 (4.5mm aperture) at the near focus.

Conclusions:

Evaluation of the three trifocal IOL models at the optical bench could show distinct peaks at the far, intermediate and near focus as well as comparable results regarding their performance.

Financial Disclosure:

... receives consulting fees, retainer, or contract payments from a company producing, developing or supplying the product or procedure presented, ... travel has been funded, fully or partially, by a competing company, ... travel has been funded, fully or partially, by a company producing, developing or supplying the product or procedure presented, ... research is funded, fully or partially, by a competing company, ... research is funded, fully or partially, by a company producing, developing or supplying the product or procedure presented, ... receives non-monetary benefits from a competing company., ... receives non-monetary benefits from a company producing, developing or supplying the product or procedure presented., ... receives consulting fees, retainer, or contract payments from a competing company

Back to previous