Posters
(results will display both Free Papers & Poster)
The number of endothelial cell five years after implantation of supricilliary micro-stent
Poster Details
First Author: M.Rau GERMANY
Co Author(s):
Abstract Details
Purpose:
To evaluate the number of the endothelial cells and the position in iridocorneal angle 5 years after implantation of supracilliary micro-stent
Setting:
Augenklinik Cam
Germany
Methods:
The study included 46 eyes which were implanted with supracilliary micro stent in 2010 -2013, 5 to 9 years ago. The number of the endothelial cells was evaluated and the position of Cypass in iridocorneal angel was observed ,
Results:
42 eyes ranged from 1900-2800,
4 eyes were between 1200 to 1400.
1 eye showed malposition of the supracilliary stent : 3 rings were visible.
In the other three eyes, there was only collar and one retention ring visible.
Conclusions:
6 % of eyes implanted with supraciliary stent showed enothelial cell loss ,2 % connected with malposition after 5 years or longer,
This group of patients is too small to allow any final assessment of the enothelial cell loss after implantation of supracilliary stent.
Financial Disclosure:
None