Meeting Highlights Information Registration Visa Letter Application Programme Overview Exhibition Virtual Exhibition Hotels Exhibition Virtual Exhibition Satellite Programme


Posters

Search Abstracts by author or title
(results will display both Free Papers & Poster)

Multicenter visual outcomes evaluation of a novel trifocal presbyopia correcting IOL: 12 months post-op results

Poster Details


First Author: T.Kohnen GERMANY

Co Author(s): A. Martinez                    

Abstract Details

Purpose:

To determine the Visual Acuity and Binocular Defocus Curve outcomes of a novel trifocal lens, the AcrySof IQ PanOptix IOL at 12 months post-operative.

Setting:

Department of Ophthalmology, Goethe-University, Frankfurt, Germany

Methods:

A prospective, single arm, non-randomized, multi-center study (17 sites) was conducted involving the bilateral implantation of AcrySof IQ PanOptix Presbyopia Correcting IOL in 151 subjects. Binocular Uncorrected Distance VA (4m), Binocular Distance Corrected intermediate VA (60cm & 80cm), Binocular Uncorrected near VA (40cm) and Binocular defocus curves were evaluated under photopic lighting conditions.

Results:

145 subjects (38% M: 62% F) with mean age of 68.9 ± 9.3 years were implanted with AcrySof-IQ PanOptix IOL bilaterally and followed up for 1 year (330-420 days). At 1-year post-op, the mean± SD (logMAR) UDVA was 0.02 ± 0.11; mean BDCIVA was 0.04 ± 0.12 and 0.08 ± 0.14 at 60cm and 80cm. Mean± SD UNVA was 0.07 ± 0.11. At 6-months follow-up, mean Defocus curve VA from 0.00D to -3.00D ranged from 0.1 to 0.0 logMAR. Average VA at distance (0.00D), intermediate (¬-1.50D) and near (-2.50D) was 0.0 ± 0.1D, 0.0 ± 0.1D and 0.1 ± 0.1D.

Conclusions:

The results from this study show very good visual performance (VA better or equal to 20/25) of the AcrySof IQ PanOptix IOL in both uncorrected and corrected conditions across the whole range of distances and defocus.

Financial Disclosure:

None

Back to Poster listing