To compare the emmetropic IOL power calculation after myopic LASIK using
OCT (Optovue), optical biometry (AL �� Scan, Nidek- Schammas PL formula) & ray tracing (Sirius, CSO)
Setting:
Mumbai Eye Care
Methods:
30 post myopic LASIK eyes were evaluated for biometry. IOL power nearest to emmetropia was noted (A constant 118.9) using above machines & compared
Results:
Mean emmetropic IOL powers using OCT, AL- scan & Sirius were 22.48D, 22.7D & 21.43. (OCT vs Al �� scan �� Not significant, OCT vs Sirius and AL- scan vs Sirius p < 0.005)
Conclusions:
OCT based IOL power calculation correlated well with optical biometry, but that using ray tracing was significantly different and more hyperopic