Posters
(results will display both Free Papers & Poster)
Comparison of corneal horizontal diameter measured by different measurements
Poster Details
First Author: Q.Chen CHINA
Co Author(s):
Abstract Details
Purpose:
To compare the accuracy of pentacam、keratograph 4、IOL Master and manualcaliper in measured corneal horizontal diameter,and refer the clinical references.
Setting:
WuhanAier Eye Hospital,Wuhan,China
Methods:
The corneal horizontal diameter was measured in 160 eyes of 80 patients using pentacam、newly developed corneal topographer、IOL Master and manualcaliper,each measurement was repeated three times ,the results were compared
Results:
The corneal horizontal diameter measured by pentacam、newly developed corneal topographer、and manualcaliper
were (11.58 ±0.29)mm、(11.48 ±0.37)mm、(11.52 ±0.36)mm,there were no significant difference between the three measurement (all P>0.05). the corneal horizontal diameter measured by IOL Master was (11.85 ±0.33)mm, there were significant difference between the IOL Master and the other three measurements(all P<0.05)
Conclusions:
Pentacam、newly developed corneal topographer、and manualcaliper can be interchanged for corneal horizontal diameter.But the corneal horizontal diameter measured by IOL Master is longer.
Financial Disclosure:
NONE