Comparison of multifocal intraocular lens power estimation using auto-keratometer and Pentacam
Poster Details
First Author: J.Han SOUTH KOREA
Co Author(s): K. Choi
Abstract Details
Purpose:
To evaluate the accuracy of multifocal intraocular lens power estimation using autokeratometer and Pentacam
Setting:
Department of Ophthalmology, Severance Hospital, Institute of Vision Research, Yonsei University College of Medicine, Seoul, Korea
Department of Ophthalmology, National Health Insurance Service Ilsan Hospital, Goyang, Gyeonggi, Korea
Methods:
99 eyes which underwent cataract surgery with Acrysof® ReSTOR® multifocal intraocular lens were analyzed. Preoperative goal diopter and actual postoperative diopter were compared. The subjects were divided into three groups based on the differences between two values, the refractive errors: -0.50 ~ +0.50 diopters(D) (group 1, n=36), less than -0.50 D (group 2, n=33), and more than +0.50 D (group 3, n=30). The subjects were also divided into another three groups according to the differences between Pentacam and autokeratometer K-values: +0.10 ~ -0.10 D (group A, n=32), less than -0.10D (group B, n=34), and more than +0.10D (group C, n=33).
Results:
Group 2 was inferior in postoperative uncorrected distance visual acuity compared to the other groups. The K-value of Pencatam in group 2 was significantly larger than the values of the other groups. Compared with other groups, group C showed significant myopic shift. With the use of Pentacam K-value, the refractive error decreased in group C.
Conclusions:
Postoperative myopic shift was confirmed when the K-value of Pentacam was larger than the value of autokeratometer. The K-value of Pentacam should be considered in case of the value is larger than the autokeratometer K for precise outcome of multifocal intraocular lens implantation.