Evaluation of a novel intraoperative toric intraocular lens alignment system using augmented reality
(results will display both Free Papers & Poster)
Session Details
Session Title: Moderated Poster Session : Cataract - Cataract Surgery Equipment/ Instruments/ Surgical Devices/ OVDs
Session Date/Time: Tuesday 08/09/2015 | 14:00-15:00
Paper Time: 14:30
Venue: Poster Village: Pod 1
First Author: : R.Varsits AUSTRIA
Co Author(s): : P. Nguyen N. Hirnschall O. Findl
Abstract Details
Purpose:
To evaluate a dedicated operating microscope with augmented reality technology to allow intra-operative toric intraocular lens (IOL) alignment without pre-operative corneal marking.
Setting:
Vienna Institute for Research in Ocular Surgery, Department of Ophthalmology, Hanusch Hospital, Vienna
Methods:
Consecutive patients who were scheduled for cataract surgery were included in this study. A novel operating microscope was used for cataract surgery that allowed superposition of a display onto the live image in one eyepiece. An automated matching of limbal vessels from pre-operative photographs taken with a modified IOL-Master 500 was used to track and follow the eye during surgery (OPMI Lumera 700 and Callisto Eye, Carl Zeiss Meditec AG, Germany). Post-operative IOL rotational alignment and rhexis size were assessed from standardised retroillumination photographs taken one hour after surgery.
Results:
In total, 50 eyes of 50 patients were included. No adverse events were reported and feasibility of the intra-operative marking was high. Deviation between the post-operative and aimed toric IOL axis was 1.91° (SD: 4.12) and absolute deviation was 3.97° (SD: 1.35).
Conclusions:
Intra-operative alignment of the IOL axis is an accurate and fast procedure resulting in precise toric IOL alignment.
Financial Interest:
NONE