Evaluation of refractive cylinder outcomes in patients undergoing cataract surgery for pre-existing astigmatism using toric IOLs or corneal incisions using the new image-guided system for surgical planning and digital alignment
(results will display both Free Papers & Poster)
Session Details
Session Title: Moderated Poster Session : Cataract - Cataract Surgery Equipment/ Instruments/ Surgical Devices/ OVDs
Session Date/Time: Tuesday 08/09/2015 | 14:00-15:00
Paper Time: 14:00
Venue: Poster Village: Pod 1
First Author: : R. Cionni FRANCE
Co Author(s): : P. Crozafon S. Slade S. Lane K. Solomon
Abstract Details
Purpose:
To evaluate residual refractive cylinder outcomes following use of a new image-guided surgical planning system with digital alignment in cataract removal and IOL implantation in patients with pre-existing astigmatism.
Setting:
Multi-center study at surgical sites in the United States and Europe.
Slade & Baker Vision Center
The Eye Institute of Utah
Carolina Eye Care
Methods:
This four-arm, open-label, prospective, observational, non-randomized, multi-center study involved 384 eyes of 277 astigmatic cataract patients. Arm 1 (70 patients; 102 eyes) involved toric IOL implantation with the Verion measurement module, vision planner and digital marker M. Arms 2a (71 patients; 95 eyes), 2b (71 patients; 87 eyes), and 2c (65 patients; 100 eyes), involved toric, monofocal, and multifocal IOL implantation, respectively, with the same system, plus digital marker L and femtosecond laser incision. Arms 2b and 2c had arcuate incisions created by the femtosecond laser. Residual refractive cylinder was measured 3 months later.
Results:
The percentage of eyes with residual refractive cylinder ≤0.5 D at 3 months was 71.6%, 74.7%, 62.1%, and 71.0% eyes in Arms 1, 2a, 2b, and 2c, respectively. The percentage of eyes with MRSE accuracy to target of less than and equal to 0.5 D at 3 months was 69.6%, 76.8%, 77%, and 82% in Arms 1, 2a, 2b, and 2c, respectively.
Conclusions:
Use of the image-based digital surgical planning and marking system shows good residual refractive cylinder results in patients undergoing toric, monofocal, or multifocal IOL implantation, with or without laser treatment.
Financial Interest:
One of the authors receives consulting fees, retainer, or contract payments from a company producing, developing or supplying the product or procedure presented