Loss of fixation in hydrophilic IOLs sulcus
(results will display both Free Papers & Poster)
Session Details
Session Title: Presented Poster Session: Cataract Surgery Complications and Management II
Session Date/Time: Saturday 05/09/2015 | 15:00-16:30
Paper Time: 15:30
Venue: Poster Village: Pod 1
First Author: : R.Quintana SPAIN
Co Author(s): : I. Cardemil M. Navarro
Abstract Details
Purpose:
To expose 3 cases with loss of stability in hidrofilic lenses fixed with prolene sutures.
Setting:
Hospital ClĂnic. Barcelona
Oftalmologia Consulting
Methods:
3 patients come to office with subluxated hidrofilic iols. We reattached the iols with prolene 10.0 suture to the sulcus passing across two eyelets of the iols. In this 3 cases the suture pass across the material with the consequence of subluxation of the iol again.
Results:
All three cases must be reoperated and re-exange the iol by a pmma iol sutured to sulcus.
Conclusions:
Hidrofilic iols will not be apropiate to suture in sulcus with prolene suture
Financial Interest:
NONE