Error induced by estimation of corneal power and effective lens position with the Mplus multifocal intraocular lens

Search Abstracts by author or title
(results will display both Free Papers & Poster)

Session Details

Session Title: Pseudophakic IOLs: Toric & Multifocal

Session Date/Time: Tuesday 08/09/2015 | 08:00-10:30

Paper Time: 09:18

Venue: Room 11

First Author: : R.PĂ©rez-CambrodĂ­ SPAIN

Co Author(s): :    D. Pinero   M. Ramon   V. Camps   V. Mateo           

Abstract Details

Purpose:

To evaluate the prediction error in intraocular lens (IOL) power calculation for a rotationally asymmetric refractive multifocal IOL and the impact on this error of the optimization of the keratometric estimation of the corneal power and the prediction of the effective lens position (ELP).

Setting:

Department of Ophthalmology (OFTALMAR), Vithas Medimar International Hospital, Alicante, Spain

Methods:

Retrospective study including a total of 25 eyes of 13 patients (age, 50 to 83 years) with previous cataract surgery with implantation of the Lentis Mplus LS-312 IOL (Oculentis GmbH). In all cases, an adjusted IOL power (PIOLadj) was calculated based on Gaussian optics using a variable keratometric index value (nkadj) for the estimation of the corneal power (Pkadj) and on a new value for ELP (ELPadj) obtained by multiple regression analysis. This PIOLadj was compared with the IOL power implanted (PIOLReal) and the value proposed by three conventional formulas (Haigis, HofferQ and Holladay I).

Results:

PIOLReal was not significantly different than PIOLadj and Holladay IOL power (p>0.05). In the Bland and Altman analysis, PIOLadj showed lower mean difference (-0.07 D) and limits of agreement (of 1.47 and -1.61 D) when compared to PIOLReal than the IOL power value obtained with the Holladay formula. Furthermore, ELPadj was significantly lower than ELP calculated with other conventional formulas (p<0.01) and was found to be dependent on axial length, anterior chamber depth and Pkadj.

Conclusions:

Refractive outcomes after cataract surgery with implantation of the multifocal IOL Lentis Mplus LS-312 can be optimized by minimizing the keratometric error and by estimating ELP using a mathematical expression dependent on anatomical factors.

Financial Interest:

NONE