Clinical results with a new 'one click' ocular biometer in 118 cataract eyes

Search Abstracts by author or title
(results will display both Free Papers & Poster)

Session Details

Session Title: Pseudophakic IOLs: Depth of Focus. Cataract Surgery Equipment

Session Date/Time: Monday 07/09/2015 | 16:30-18:30

Paper Time: 18:02

Venue: Room 16

First Author: : M.Mrochen SWITZERLAND

Co Author(s): :    G. Kervick   M. Galligan   D. Bogusevschi   A. Nolan   A. Cummings        

Abstract Details

Purpose:

A new ocular “one-click” biometer (MirriconTM, ClearSight Innovations, Dublin, Ireland) providing ocular parameter by high-resolution noncontact measurements. The study evaluates the validity and repeatability of these measurements compared with current clinical instrumentation.

Setting:

Mater Private Hospital, City Gate, Mahon, Cork, Ireland; ClearSight Innovations, Dublin, Ireland; IROC Science, Zurich, Switzerland

Methods:

Measurements were taken with the Mirricon, OB820 (Alcon WaveLight, Erlangen, Germany) and the Oculyzer (Pentacam, Oculus, Wetzlar, Germany) on 118 eyes of 98 patients aged from 48 to 89 years listed for cataract surgery. The LOCS scaling ranged from I to IV. Correlation of anterior radius of curvature and anterior chamber depth was compared among all 3 devices. Axial length and lens thickness between Mirricon and OB820 as such metrics are not available from the Oculyzer. The posterior corneal astigmatism was compared only to the Oculyzer as such metric are not available from the OB820.

Results:

The correlation coefficients for all metrics were found to be significant and ranging from R = 0.81 for Posterior Corneal Radius to 0.99 for Axial Length. All eyes could be measured with the Mirricon, 7 eyes were not measureable with the OB820 and all eye could be measured with the Oculyzer.

Conclusions:

Mirricon combines the information provided from OB820 and Oculyzer in one device. The clinical results demonstrate equivalent measurement results. Mirrion could successfully measure more eyes than the OB820.

Financial Interest:

One of the authors has significant investment interest in a company producing, developing or supplying product or procedure presented, One of the authors is employed by a competing company