Pupil size alterations induced by photodisruption during femtosecond laser-assisted cataract surgery

Search Abstracts by author or title
(results will display both Free Papers & Poster)

Session Details

Session Title: Interactive Free Paper Session: FLACS

Session Date/Time: Sunday 06/09/2015 | 08:30-10:00

Paper Time: 09:08

Venue: Room 16

First Author: : M.Choi SOUTH KOREA

Co Author(s): :    J. Jun   S. Chung   C. Joo              

Abstract Details

Purpose:

To measure miosis after femtosecond laser pretreatment of cataract surgery and determine correlative factors.

Setting:

Department of Ophthalmology and Visual Science, Seoul St. Mary's Hospital, College of Medicine, and Catholic University of Korea, Seoul, South Korea. Cross-sectional study.

Methods:

Images extracted from surgical videos of femtosecond laser pretreatment and phacoemulsification were used to measure the pupil area. Quantitative analysis of pupil constriction was performed by comparing consecutive images obtained at the initiation of each procedure. Potential factors related to pupil constriction were examined, including laser parameters and anatomic measurements

Results:

The study enrolled 56 eyes. The mean pupil area decreased by 29.7% during the time it took to shift between procedures (shifting time). The total laser treatment duration was highly correlated with the reduction in the pupil area (r Z 0.433, P Z .001). The amount of intraoperative miosis was significantly correlated with the duration of lens fragmentation by the femtosecond laser (rZ0.416, PZ.001) and of primary incision creation (rZ0.289, PZ.031). Pupil constriction was correlated with patient age (r Z 0.398, P Z .002) and the laser capsulotomy–pupil margin distance (r Z 0.395, P Z .003), but not with suction-on time (r Z 0.012, P Z .930) or shifting time (r Z 0.091, P Z .506).

Conclusions:

The pupil area decreased significantly after femtosecond laser pretreatment of cataract surgery. Intraoperative miosis was most significantly correlated with laser pretreatment duration and patient age.

Financial Interest:

NONE