Xen Gel Stent vs Preserflo MicroShunt in open-angle glaucoma (OAG)
Free Paper Details
First Author: M.Mackert GERMANY
Co Author(s): S. Freissinger M. Nobl E. Vounotrypidis S. Priglinger
Abstract Details
Purpose:
Comparison of two novel glaucoma drainage devices with subconjunctival filtration (Xen and Microshunt) for OAG regarding effectiveness and safety.
Setting:
This single-centre, retrospective, comparative cohort study was performed at the Ludwig-Maximilians-University eye hospital, Munich, Germany.
Methods:
Primary endpoint was the intraocular pressure (IOP) reduction 3-4 months after surgery. Secondary endpoints were the reduction of glaucoma medications, intraoperative and postoperative adverse events and needling rate. For statistical analysis the Wilcoxon test (p less-than 0.05) and the multivariate analysis of variance (MANOVA) was performed using SPSS.
Results:
Seventy-eight eyes of 78 patients received either Xen (n=52) or Microshunt (n=26) to control IOP in OAG. IOP reduction after 3-4 months postoperative was statistically significant (p less-than 0.001) in both groups (mean -5,9mmHg versus -9mmHg) with a higher IOP lowering effect in the Microshunt group (p=0.031). A statistically significant reduction in glaucoma medications was achieved in both groups (p less-than 0.001) with an advantage in the Microshunt group (p=0.012). No severe adverse events were observed in any group. The needling rate was 23% in the Xen group and 17% in the Microshunt group.
Conclusions:
Xen and Microshunt are effective glaucoma draining devices significantly reducing both IOP and glaucoma medications in OAG. Overall, Microshunt seems to be a more effective treatment option in comparison with Xen in patients with OAG.