Information Registration Programme Overview Satellite Education Programme Virtual Exhibitor Listing


ePoster

All ePoster presentations will be available to view anytime throughout the Meeting in an Online Library

Search Abstracts by author or title

Title:

Comparison of TrabEx+ and trabectome-mediated irrigating goniectomy


Poster Details

First Author: N. Loewen GERMANY

Co Author(s):    Y. Al Yousef   A. Strzalkowska                 

Abstract Details

Purpose:

To determine whether excisional irrigating goniectomy with the TrabEx+ is as safe and effective as plasma-mediated ablation with the Trabectome.

Setting:

Prospective interventional cohort. University Eye Clinic, W�rzburg, Germany.

Methods:

Comparison of results in 90 TrabEx+ patients to Trabectome. The TrabEx+ and Trabectome (Microsurgical Technology, Bellevue, WA, US) both remove the trabecular meshwork (TM), using different approaches. The TrabEx+ uses a double blade to excise while the Trabectome molecularizes the TM using plasma. 98 TrabEx patients will be enrolled and compared to existing Trabectome data.

Results:

To date, 90 TrabEx+ patients have been enrolled and completed the 3-month follow-up visit. Tmax was 28.8�11 mmHg. The preoperative IOP of 18.0�7.0 reduced to 13.1�3.2 (day 30) and 15.1�5.6 mmHg (day 90). In the 165 Trabectome patients, IOP dropped from 22.3�5.6 to 16.7�5.6 (day 30) and 16.0�3.9 mmHg (day 90). Medications in TrabEx+ declined from 2.8�1.8 to 2.4�1.4 (day 30) and 2.7�1.7 (day 90) compared to from 2.7�1.1 to 2.4�1.5 (day 30) and 2.3�1.5 (day 90) in Trabectome. Both TrabEx+ and Trabectome achieved a significant IOP reduction without intergroup differences (p<0.05). No vision-threatening complications occurred in either group.

Conclusions:

TrabEx+ and Trabectome lowered IOP to the same extent and in a safe manner during the time observed.

Financial Disclosure:

None




Back to Poster listing