Comparison of central corneal thickness measurements obtained by Pentacam HR and Ultrasound Pachymeter in LASEK candidates

Poster Details


First Author: V.Damanpak IRAN

Co Author(s): B. Samet   A. Dustdar                 

Abstract Details

Purpose:

to compare central corneal thickness measurements obtained by Pentacam HR scheimpflug system((Oculus, Weltzar, Germany) and Ultrasound Pachymeter(Tomey SP100) in LASEK refractive surgery candidates

Setting:

Iran,Tehran, Vanak Eye Center

Methods:

In this prospective study ,central corneal thickness of 264 eyes of 132 LASEK candidates were measured with Pentacam HR and ultrasound Ultrasound Pachymeter then results were analyzed with paired T test ,regression analysis and Bland-altman plots.

Results:

The mean central corneal thickness values with Pentacam HR and Ultrasound Pachymeter were 539.87 ± 32.29 and 540±32.67 respectively.there was a significant linear correlation between Pentacam HR and Ultrasound Pachymetry measurements(r=0.95 , p<0.001). the mean difference between Pentacam HR and Ultrasound Pachymeter was not statistically significant(paired t test : p=0.24)

Conclusions:

the central corneal thickness values obtained by these two modalities are highly correlated with together and Pentacam can be alternative to Ultrasound Pachymetry in measuring central corneal thickness

Financial Disclosure:

None