Effect of intrastromal corneal ring segment depth on visual, refractive and topographic improvement
(results will display both Free Papers & Poster)
Session Details
Session Title: Cornea
Session Date/Time: Friday 09/02/2018 | 08:30-10:00
Paper Time: 09:21
Venue: Annex A
First Author: R.Carneiro de Freitas PORTUGAL
Co Author(s): T. Monteiro N. Franqueira F. Faria Correia F. Vaz
Abstract Details
Purpose:
To analyze the effect of different ICRS depths on visual, refractive and topographic parameters
Setting:
Hospital de Braga, Portugal
Methods:
After manual implantation, ICRS depth was measured 6 months after the procedure using Swept-Source AS-OCT (Casia SS-1000, Tomey), and patients were divided into 2 groups: group 1 included patients in which the ICRS was located ≤30µm away from the intended depth, and group 2 included patients in which the ICRS was > 30 µm away from the intended depth of implantation. Visual, refractive and topographic parameters were measured and vectorial astigmatism analysis (Thibos method) was performed and compared between both groups at 6 months follow-up.
Results:
Group 1 included 36 patients and Group 2 included 63 patients. No statistical difference was found between both groups at baseline in any of the parameters evaluated. Both UDVA and CDVA improved in both groups (p<0.01), and no difference was found between both groups (UDVA p=0.79; CDVA p=0.86). No statistical difference was seen between the 2 groups regarding refractive sphere, refractive cylinder and topographic cylinder at 6M follow-up. In Group 1 no change was observed in J0, J45 and B, while a change in the vector magnitude could be seen in Group 2 at 6 months follow-up.
Conclusions:
Different depths of ICRS implantation in relation to the primarily intended depth did not show a significant effect on visual, refractive and topographic parameters.
Financial Disclosure:
None